PROCESS TURBIDITY SENSOR
|
![]() |
The Monitek® MoniTurb sensor, in connection with a Monitek® Messenger transmitter, measures turbidity in liquids. |
The system can be used for inspection, control and optimization of industrial processes:
Drinking water
Waste water
Chemical industry
Paper industry
Biotechnology
Beverages
Ultra filtrate
Filtrate
Pre-filter
Condensate
Process water
Water in oil / oil in water
DESCRIPTION
The MoniTurb-F turbidity sensor uses the principle of 12° forward scattered light to detect suspended particles in liquids. The ratio of scattered light to direct light assures highly reliable and repeatable measurement results. The 12° scattered light measuring result is dependent on particle size and can be correlated to mass concentration. The sensor can be installed into almost any type of pipe. Process connection, pressure, temperature, gasket material, etc. will be application specific. This product is also available as the MoniTurb-S, which is a 90° side scatter turbidity sensor, and the Moniturb-FS carrying out both forward scatter and side scatter measurements.
The optional feature for hazardous area, Class I Div. 1, Zone I or Zone II, allows for wide range of applications. A maximum cable length of 100 m (approx. 330 feet) allows the converter to be located outside the hazardous area.
FEATURES
MoniTurb-F uses 12° Forward Scattered Light
90° Side Scatter option available: MoniTurb-S
Combined Forward / Side Scatter option: MoniTurb-FS
Calibration interval: 12 months
Color compensation
Cleaning Jets (optional)
Protection: NEMA 4X / IP65
Process connections: DIN, ANSI, SMS, NPT, APV, TH, Tri-Clover.
Optional hazardous area: Class 1 Div. 1, Ex-zone I and Ex-zone II
Compatible to converter type: Monitek® Messenger
SPECIFICATIONS
Line sizes: 10 – 100 mm (1/2” – 4”)
Process pressure: Max. 220 psig at ambient temp./15 bar
Process temperature: Max. 140° C
Gasket material: Application specific
Sensor material: 316 L SS / 1.4404 (other materials on request)
Optical path length: Depends on pipe size
Cable length: Max. 100 m (330 ft)
Measurement range: 0 - 10 ppm to 0 – 500 ppm
Detector system: Silicone diodes
Accuracy: +/- 2%
Reproducibility: +/- 1%
Click here for the complete brochure on the Monitek Moniturb Turbidity Sensor,
Monispec A, and Monispec AD Sensors (PDF - 665K)
Click here to view industry application notes for Monitek
Click here for the Application Data Collection Sheet (PDF - 33K)
![]() |
![]() |